Fuses & power settings – Carbolite VMF Series User Manual
Page 15
AAF, GSM, VMF
MF06 3.32
15
8.0 F
USES
&
P
OWER
S
ETTINGS
8.1 Fuses
F1-F3: Refer to the circuit diagrams.
F1
Internal
supply
fuses
Fitted if supply cable fitted.
Fitted on board to some types of EMC filter.
on-board and up to 16 Amps:
32mm x 6mm type F
other: GEC Safeclip
F2
Auxiliary
circuit
fuses
Fitted on board to some types of EMC filter.
May be omitted up to 25Amp/phase supply
rating.
2 Amps glass type F
On board: 20mm x 5mm
Other: 32mm x 6mm
F3
Heat Light
fuses
May be omitted up to 25 Amp/phase supply
rating.
2 Amps glass type F
32mm x 6mm
Customer
fuses
Required if no supply cable fitted.
Recommended if cable fitted.
See rating label for amperage;
see table below for fuse rating.
Model
Phases
Volts
Supply Fuse
rating
AAF 11/3
1-phase
200-240
10A
1-phase
110
20A
AAF 11/7
1-phase
1-phase
1-phase
2-phase+N
220-240
200-208
110-120
380/220-415/240
20A
20A
40A
10A per phase
AAF 11/18
1-phase
220-240
30A
1-phase
200-208
40A
3-phase+N
380/220-415/240
12.5A per phase
3-phase delta
220-240
20A per phase
3-phase delta
200-208
25A per phase
AAF 12/3
1 phase
200 - 240
10A
AAF 12/7
1-phase
220–240
25A
AAF 12/18
1-phase
220-240
30A
1-phase
200-208
40A
3-phase+N
380/220-415/240
12.5A per phase
3-phase delta
220-240
20A per phase
3-phase delta
200-208
25A per phase
GSM 11/8
1-phase
1-phase
1-phase
220-240
200-208
110-120
12.5A
15A or 16A
25A
VMF 10/6
1-phase
220-240
12.5A
8.2 Power Settings
The settings of the power limit parameter
OP.Hi
are as follows.
model volts:
110
200
208
220, 380
230, 400
240, 415
254, 440
AAF 11/3
87
100
98
87
80
73
65
AAF 11/18
100
100
100
100
94
83
AAF 12/3
100
95
85
78
71
AAF 12/7
100
100
100
AAF 12/18
100
100
100
100
94
83
all other models
100
100
100
100
100
89