Maintenance – Carbolite HRF 7/45 User Manual
Page 5
HRF
MLF05-1.23
5
Avoid burns. Carbolite can supply tongs, face masks, and heat resistant gloves. Before you remove
a hot object from the furnace make sure you have a safe place to put it down.
4.0 M
AINTENANCE
4.1 General Maintenance
No routine maintenance is required other than removal of soot deposits mentioned in 3.2 and the
occasional replacement of consumable items.
The furnace outer surface may be cleaned with a damp cloth. Do not allow water to enter the
interior of the case or chamber. Do not clean with organic solvents.
4.2 Calibration
After prolonged use the controller and/or thermocouple could require recalibration. This would be
important for processes which require accurate temperature readings or which use the furnace
close to its maximum temperature. A quick check using an independent thermocouple and
temperature indicator should be made from time to time to determine whether full calibration is
required. Carbolite can supply these items.
Depending on the controller, the controller manual may contain calibration instructions.
4.3 After Sales Service
Carbolite’s service division (Carbolite Engineering Services) has a team of Service Engineers
capable of repair, calibration and preventive maintenance of furnace and oven products at our
customers’ premises throughout the world. We also sell spares by mail order. A telephone call or
fax often enables a fault to be diagnosed and the necessary spare part despatched.
Each furnace has its own record card at Carbolite. In all correspondence please quote the serial
number, model type and voltage given on the rating label of the furnace. The serial number and
model type are also given on the front of this booklet when supplied with a furnace.
To contact Carbolite Engineering Services or Carbolite see the back page of this manual.
4.4 Recommended Spares Kits
Carbolite can supply individual spares, or a kit of the items most likely to be required. Ordering a
kit in advance can save time in the event of a breakdown. When ordering spares please quote the
model details as requested above.
4.5 Power Adjustment
The furnace control system incorporates electronic power limiting, but in these models the power
limit is set to 100% (but see section 8.2 for any exceptions).
Occasionally the power limit is set to zero to permit demonstration of the controls without the
heating elements taking power. In this case the power limit may be reset to its standard value to re-
enable heating.
The power limit may not be accessible to the operator, in which case, if it is necessary to alter the
limit, Carbolite can assist with instructions.