7.3 3-phase “universal” wiring

Fuses F1 are always present in this wiring design.

Fuses F2 could be absent in some circumstances, if the circuit does not exceed 10A.

Fuses F3 are present if the circuit exceeds 25A, but otherwise are usually absent.












Instrument Switch

coil

temperature

controller

overtemp.

controller

thermocouple

SSR

heat

on

F3

F1

F2

F3

SSR

SSR

F3

heat

on

F3

e

l

e

m

e
n

t

(s)

heat

on

F3

F3

if fitted

contactors

or relays

Filters

e

l

e

m

e
n

t

(s)

e

l

e

m

e
n

t

(s)

coil

L

N

L

N

L

N

thermocouple

A model made to this diagram can be
converted by the customer between the
following supply voltages:

between 3-phase + neutral in the range

380/220V – 415/240V

and

3-phase delta in the range
220V – 240V

and

single phase in the range
220V – 240V

3-phase + neutral

Circuit no. 1

N

L

N

L

Circuit no. 2

Circuit no. 3

N

L3

L2

L1

Supply
Terminals

N

L

Fuses or
EMC filters

PE

Not
used

3-phase delta

Circuit no. 1

N

L

N

L

Circuit no. 2

Circuit no. 3

N

L3

L2

L1

Supply
Terminals

N

L

Fuses or
EMC filters

PE

1-phase

Circuit no. 1

N

L

N

L

Circuit no. 2

Circuit no. 3

N

L3

L2

L1

Supply
Terminals

N

L

Fuses or
EMC filters

PE

Not
used