Installation – Carbolite AX120 User Manual
Page 3

Apex Series Ovens
MF42 1.10
3
2.0 I
NSTALLATION
2.1 Unpacking & Handling
Remove the shelves, runners and hangers from the packaging before installing the equipment.
Lift the unit by its base. The door should not be used to support the equipment when moving it.
Use two people to carry the oven where possible. Remove any packing material from the inner
chamber before use.
2.2 Siting & Setting Up
Place the oven on a level surface. If the overtemperature protection option is not fitted, ensure that
the unit can be directly observed.
Ensure that there is at least 50mm of free space behind the oven and 25mm at the sides. There are
vents in the back that must not be obstructed.
Ensure that the oven is placed in such a way that it can be quickly switched off or disconnected
from the electrical supply - see below.
The two shelf hangers should be fitted into the slots provide on the left
hand side of the chamber. The vertical parts of the runners should be
inserted into the holes in the hangers (on the left) or the fan cover (on
the right), front and back simultaneously; the runner should then be
rotated in a downwards direction to secure it in place. The shelves slide
onto the runners such that the spurs on the lower side of the shelf are
under the runner at the back, preventing the shelf from tilting forwards
when partially withdrawn.
Plan to use the shelves, not the oven floor (see section 3.6).
2.3 Electrical Connections
Connection by a qualified electrician is recommended.
The Apex ovens are made only for single phase A.C. supply, which may be Live to Neutral non-
reversible (polarised), Live to Neutral reversible (non-polarised) or Live to Live. Check the oven
rating label before connection. The supply voltage should agree with the voltage on the label, and
the supply capacity should be sufficient for the amperage on the label.
The supply should be fused at the next size equal to or higher than the amperage on the label. The
fuse ratings are given in section 7.3 of this manual. Internal supply fuses are fitted in these models,
but customer fusing is also recommended.
The oven is fitted with a supply cable. This may be fitted with a line plug or wired directly to an
isolator. Ensure that the unit can be quickly disconnected from the supply.
The supply MUST incorporate an earth (ground).
CONNECTION DETAILS
supply type
Supply
Terminal label
Cable colour
Live-Neutral
Reversible or Live-Live
1-phase
L
Brown
To live
to either power conductor
N
Blue
To neutral
to the other power conductor
PE
Green/Yellow
To earth (ground)
to earth (ground)