Apex Series Ovens


10

MF42 1.10

6.0 F

AULT

A

NALYSIS

A. Oven Does Not Heat Up

1.

The temperature
controller is OFF

 No power from the

supply

 Check the fuses in the supply

line

 There are no lights

glowing on the
controller

 The controller may be faulty or

not receiving a supply due to a
faulty switch or a wiring fault

2.

The temperature
controller is ON

 The controller shows a

very high temperature
or a code such as EEE or
---

 The temperature sensor has

broken or has a wiring fault

 The controller shows a

low temperature

 The SSR could be failing to

switch on due to internal
failure, faulty logic wiring
form the controller, or faulty
controller

B. Oven Overheats

1.

Oven only heats up
when the instrument
switch is ON

 The controller shows a

very high temperature

 The controller is faulty

 The controller shows a

low temperature

 The temperature sensor may

not be positioned in the oven
correctly

 The controller may be faulty

2.

Oven heats up when
the instrument switch
is OFF

 The SSR has failed

“ON”

 Check for an accidental wiring

fault that could have
overloaded the SSR

C. Process Timer

Oven will not heat

 Temperature controller

does not light up

 Process timer has not been

reset

 Resetting the timer has

no effect

 Process timer is faulty (relay

stuck open)

Oven continues to heat
at end of process time

 Process timer OUT light

is continuously on

 Process timer is faulty (relay

stuck closed)

 OUT light flashing or

off

 not a process timer fault

This manual is related to the following products: