5 start and exit software, 1 start software, 2 exit software – Eppendorf BioPhotometer Data Transfer User Manual
Page 14: 6 gather data, 1 optional settings prior to saving data, Start and exit software 6.5.1, Start software, Exit software, Gather data 6.6.1, Optional settings prior to saving data

6 Operation
14
BioPhotometer Data Transfer Software — Operating Manual
6.5
Start and exit software
6.5.1
Start software
Start the BioPhotometer Data Transfer Software via the shortcut
on the desktop.
You are directed to the area Data Output of the Work area.
6.5.2
Exit software
1. Select Exit in the Menu File.
2. Confirm the following query with Yes.
6.6
Gather data
There are two possibilities to transfer data with the BioPhotometer Data Transfer Software from
BioPhotometer plus to the PC.
1. Gather data simultaneously for measurement at BioPhotometer plus.
2. Transfer data from the results memory of BioPhotometer plus to the PC.
Requirement
•
The BioPhotometer plus is connected to the PC via the provided null modem cable.
•
The BioPhotometer plus is switched on and LANGUAGE ENGLISH or LANGUAGE U.S.ENGL is
set as the menu language. With a different language setting it may result in faulty data
transfer.
•
The software is started.
•
The software has established a connection to BioPhotometer plus if the following example
message is displayed in the Data Output in the Work area:
Connected device: BioPhotometer plus 6132 found at COM1.
6.6.1
Optional settings prior to saving data
Create archive file
Creating a new archive file is described separately(see Create archive file on page 24).
Measurement ID: assign identification code
Identification codes of up to 20 digits can be assigned to a measurement series or also individual
measurements:
1. Click on the icon
in the function strip.
An input window opens:
2. Enter the identification code
Hint!
The preset name of the archive file is ResultData. You can use it or prior to recording the first
measurement data create a new archive file.
6
Operation