4 Maintenance
The
in situ
Adapter does not require regular maintenance.
Contamination can be removed using water or a mild laboratory cleaning
agent. The covering hood should be wiped using a moist cloth only in
order to prevent water or other liquids from entering the lateral ventilation
slits of the double chamber.
Do not allow any parts of the
in situ
Adapter to come into contact with
organic solvents or aggressive chemical. Do not clean the transparent
plastic parts using alcohol or acetone.
5 Technical data
Type of instrument:
in situ
Adapter for Mastercycler
or Mastercycler
gradient
Sample capacity:
Max. 4 glass slides of 76 x 26 mm
(3 x 1 inch)
Temperature range:
4 to 99 °C
Temperature distribution
40 °C: ± 0,6 °C
across the adapter:
72 °C: ± 0,6 °C
95 °C: ± 1,0 °C
Temperature-control speed:
Heating rate
≥
1,0 °C/s,
measured on the adapter
Cooling rate
≥
0,7 °C/s,
measured on the adapter
Dimensions of the adapter:
Metal carrier:
Width: 112 mm
Depth:
79 mm
Height:
9 mm
Covering hood: Width: 148 mm
Depth: 105 mm
Height: 31 mm
Weight:
Metal carrier: approx. 60 g
Conditions of use:
0 to 105 °C,
25 to 100 % relative humidity
Storage conditions:
–20 to +70 °C,
25 to 85 % relative humidity