100

100 – 1000 and 500 – 2500 μL (Fig. 6)

Push the control button all the way down and pull off the
ejector sleeve (13).
Piston seal (4)
– Unscrew pipette at the central junction.
– Discard piston seal (4) and replace.
Ejector seal (10)
– Unscrew lower part of the pipette.
– Discard seal (10) with O-ring.
– Insert new seal (see "Inserting the seals").
Tighten the tip cone (see IV of this part).
The tip cone (11) is also available complete with ejector
seal.

A

B

C

B

A

C

C

A

B

4

13

11

10

Eppendorf Reference® – Part B – Maintenance