NOTICE
NOTICE! Damage to device from missing pipette tips.
Only use the Reference with attached pipette tips.
NOTICE
NOTICE! Carry-over, contamination and incorrect dispensing results
due to the repeated use of pipette tips.
The pipette tips are for single use only. Prolonged use can have a negative
impact on dispensing tasks.
Use the pipette tips only once.
Do not use autoclaved ep Dualfilter T.I.P.S. for dispensing.
NOTICE
NOTICE! Incorrect dispensing volume with special liquids and due to
temperature differences.
Solutions which differ greatly from water in terms of their physical data, or
temperature differences between the pipette, pipette tip and liquid can result
in incorrect dispensing volumes.
Avoid temperature differences between pipette, pipette tip and liquid.
Ensure that the temperature remains constant between 20 and 25 °C,
±0,5 °C.
Check the dispensing volume and readjust the pipette in case of
deviations.
NOTICE
NOTICE! Damage to device due to penetration of liquids.
Do not deposit the Reference with filled pipette tip.
Only use pipettes in the Reference-Serie when a pipette tip has been
attached.
Do not allow any liquids to penetrate the inside of the housing.
The intended use is described in a separate document enclosed in the
package.