beautypg.com

3 reference 2 fixed, Reference 2 fixed – Eppendorf Pipette Reference 2 User Manual

Page 35

background image

Technical data

Eppendorf Reference

®

2

English (EN)

35

7.3

Reference 2 fixed

Reference 2 fixed volume

Model

Test tip
epT.I.P.S. color code
Volume range
Length

Eppendorf

AG error limits

Error

Systematic error

Random error

±

%

±

µL

±

%

±

µL

1

μL

Dark

gray

0.1

μL – 10

μL

34 mm

±2.5

±0.025 ±1.8

±0.018

2

μL

±2.0

±0.04

±1.2

±0.024

5

μL

Medium

gray

0.1

μL – 20

μL

40 mm

±1.2

±0.06

±0.6

±0.03

10 μL

±1.0

±0.1

±0.5

±0.05

20

μL

Light

gray

0.5

μL – 20 μL L

46 mm

±0.8

±0.16

±0.3

±0.06

10 μL

Yellow
2

μL

200

μL

53 mm

±1.2 ±0.12

±0.6

±0.06

20

μL

±1.0

±0.2

±0.3

±0.06

25

μL

±1.0

±0.25

±0.3

±0.075

50

μL

±0.7

±0.35

±0.3

±0.15

100

μL

±0.6

±0.6

±0.2

±0.2

200

μL

±0.6

±1.2

±0.2

±0.4

200

μL

Blue
50

μL – 1000

μL

71 mm

±0.6

±1.2

±0.2

±0.4

250

μL

±0.6

±1.5

±0.2

±0.5

500

μL

±0.6

±3.0

±0.2

±1.0

1000

μL

±0.6

±6.0

±0.2

±2.0

Fixed-volume Reference 2

Model

Test tip
epT.I.P.S. color code
Volume range
Length

Eppendorf

AG error limits

Error

Systematic error

Random error

±

%

±

mL

±

%

±

mL

2.0

mL

Red
0.5

mL – 2.5

mL

115 mm

±0.6

±0.012

±0.2

±0.004

2.5

mL

±0.6

±0.015

±0.2

±0.005