2 reference 2 multi-channel variable, Reference 2 multi-channel variable – Eppendorf Pipette Reference 2 User Manual
Page 34
Technical data
Eppendorf Reference
®
2
English (EN)
34
7.2
Reference 2 Multi-channel variable
Reference 2 single-channel variable
Model
Test tip
epT.I.P.S. color code
Volume range
Length
Testing volume
Eppendorf
AG error limits
Error
Systematic error
Random error
±
%
±
mL
±
%
±
mL
0.25
mL – 2.5
mL
Increment:
0.002
mL
Red
0.25
mL – 2.5
mL
115
mm
0.25
mL
±4.8
±0.012
±1.2
±0.003
1.25
mL
±0.8
±0.010
±0.2 ±0.0025
2.5
mL
±0.6
±0.015
±0.2
±0.005
0.5
mL – 5
mL
Increment:
0.005
mL
Purple
0.1
mL – 5
mL
120 mm
0.5
mL
±2.4
±0.012 ±0.6
±0.003
2.5
mL
±1.2
±0.030
±0.25 ±0.006
5.0
mL
±0.6
±0.030
±0.15
±0.0075
1
mL – 10
mL
Increment:
0.01
mL
Turquoise
1
mL – 10
mL
165 mm
1.0
mL
±3.0
±0.030
±0.6
±0.006
5.0
mL
±0.8
±0.040
±0.2
±0.010
10.0
mL
±0.6
±0.060
±0.15
±0.015
Reference 2 Variable multi-channel (8-channel/12-channel)
Model
Test tip
epT.I.P.S. color code
Volume range
Length
Testing volume
Eppendorf
AG error limits
Error
Systematic error
Random error
±
%
±
µL
±
%
±
µL
0.5
μL – 10
μL
Medium
gray
0.1
μL – 20
μL
40
mm
0.5
μL
±12.0
±0.06 ±8.0 ±0.04
1
μL
±8.0 ±0.08
±5.0 ±0.05
5
μL
±4.0 ±0.2 ±2.0 ±0.1
10
μL
±2.0 ±0.2 ±1.0 ±0.1
10
μL – 100
μL Yellow
2
μL
–
200
μL
53
mm
10
μL
±3.0 ±0.3 ±2.0 ±0.2
50
μL
±1.0 ±0.5 ±0.8 ±0.4
100
μL
±0.8 ±0.8 ±0.3 ±0.3
30
μL – 300
μL Orange
20 μL – 300 μL
55
mm
30
μL
±3.0 ±0.9 ±1.0 ±0.3
150
μL
±1.0 ±1.5 ±0.5 ±0.75
300
μL
±0.6 ±1.8 ±0.3 ±0.9