beautypg.com

Eppendorf Pipette Reference 2 User Manual

Page 14

background image

Safety
Eppendorf Reference

®

2

English (EN)

14

NOTICE! Damage to device from missing pipette tips.

Use the Reference 2 with fitted pipette tips only.

When using pipette tips without filters, use 2-10

mL pipettes with protective sleeves and

an inserted protection filter.

NOTICE! Carry-over, contamination and incorrect dispensing results due to the incorrect
use of pipette tips.
The pipette tips are for single use only. Prolonged use can have a negative impact on
dispensing tasks.

Only use the pipette tips once.

Do not use autoclaved ep Dualfilter T.I.P.S. for dispensing.

NOTICE! Incorrect dispensing volume due to differences in temperature and for special
liquids.
Temperature differences between the pipette, pipette tips and liquid; or solutions with
physical data that is very different from that of water; may result in incorrect dispensing
volumes.

Avoid temperature differences the between pipette, pipette tip and liquid.

Ensure that the temperature is between 20

°C and 27

°C and remains constant at ±0.5

°C.

Check the dispensing volume and make sure that you can affirm all the questions listed in
the general information.

NOTICE! Damage to the device due to penetration of liquids.

Do not put the pipette down when the pipette tip is filled.

Do not allow any liquids to penetrate the inside of the housing.

If liquid has entered the inside of the housing, the inner parts may only be repaired by
Eppendorf AG service partners. Contact your local sales office before returning any
devices.