Figure 7: Chamber Alarm System Flow Chart


15 MINUTES

ALARM DELAY

TIMEOUT

COMPLETED

ARE TEMPERATURES AND CO2

LEVELS WITHIN +/- 0.1 OF SET

POINT?

CHANGE

PROGRAMMED

VALUES

OPEN INNER DOOR

DISARM ALARMS

CLOSE INNER

DOOR

15 MINUTES

ALARM DELAY

TIMEOUT STARTED

IS TEMPERATURE OR CO2 LEVEL

MORE THAN +/- 0.5 OUTSIDE

PROGRAMMED POINTS?

CANCEL ALARM BY

PRESSING ANY KEY

ALARMS

CANCELLED

VISUAL AND AUDIBLE

ALARMS CANCELLED,

MESSAGE LEFT ON SCREEN

CANCEL

MESSAGE BY

PRESSING ANY

KEY

NO

YES

YES

SWITCH ON

ARM ALARMS

VISIBLE AND AUDIBLE

ALARM - MESSAGE ON

SCREEN

HAVE CONDITIONS

RECOVERED TO

WITHIN +/- 0.1?

YES

NO

NO


3.8.1

Temperature Sensor System Alarms

There are four temperature sensors: two in the door and two in the chamber. If
any of these sensors should fail, the following message will appear:

°C FAIL

.


Because the incubator can no longer control temperature properly without the
failed sensor, the heating will switch off and the incubator will cool down to
room temperature.