Warning – Eppendorf S Series 48S CO2 Incubator User Manual
Page 16
16
CO48S-0050
Galaxy S Series 48 S Personal Size CO
2
Incubator
User’s Guide
Connect the CO
2
gas supply:
WARNING!
Slightly increased levels of CO
2
may be found in and around the operating
area of the CO
2
incubator. Over time, this can have adverse effects on
those exposed to such an environment. Users working in environments
with elevated levels of CO
2
should take all appropriate precautions to
protect their breathing.
1. Connect the incubator to the CO
2
supply using the ~1/4-inch or 6mm plastic tubing
(with installed HEPA filter) by attaching the tubing from the two-stage regulator (or
in-line regulator) to the matching CO
2
inlet on the rear of the incubator.
NOTE:
It is highly recommended that an in-line regulator be used at the
incubator’s gas inlet(s).
For proper incubator operation, CO2 gas pressure must not exceed 5 psi
(0.35 bar).
2. Use the tubing clips provided to eliminate CO
2
leaks.
Install the HEPA AutoZero filter:
1. Press the AutoZero HEPA filter gently into the white plastic filter socket at the top
of the rear panel of the incubator (see Figure 4).
Figure 4: AutoZero HEPA Filter & Fitting