beautypg.com

Warning – Eppendorf S Series 48S CO2 Incubator User Manual

Page 16

background image

16

CO48S-0050

Š Galaxy S Series 48 S Personal Size CO

2

Incubator

User’s Guide

Connect the CO

2

gas supply:

WARNING!

Slightly increased levels of CO

2

may be found in and around the operating

area of the CO

2

incubator. Over time, this can have adverse effects on

those exposed to such an environment. Users working in environments
with elevated levels of CO

2

should take all appropriate precautions to

protect their breathing.

1. Connect the incubator to the CO

2

supply using the ~1/4-inch or 6mm plastic tubing

(with installed HEPA filter) by attaching the tubing from the two-stage regulator (or
in-line regulator) to the matching CO

2

inlet on the rear of the incubator.

NOTE:

It is highly recommended that an in-line regulator be used at the
incubator’s gas inlet(s).

For proper incubator operation, CO2 gas pressure must not exceed 5 psi
(0.35 bar).

2. Use the tubing clips provided to eliminate CO

2

leaks.


Install the HEPA AutoZero filter:

1. Press the AutoZero HEPA filter gently into the white plastic filter socket at the top

of the rear panel of the incubator (see Figure 4).

Figure 4: AutoZero HEPA Filter & Fitting