Eppendorf S Series 48S CO2 Incubator User Manual

Page 24

background image

24

CO48S-0050

Š Galaxy S Series 48 S Personal Size CO

2

Incubator

User’s Guide

If temperature and/or CO

2

levels deviate more than the programmed amount, the

display flashes, the audible alarm sounds and a message appears on the screen. You
can acknowledge (and cancel) the alarm by pressing any key.

When the inner door (if fitted) is opened, the Alarm System is disabled. When you
close the outer door, the preset Alarm Arming Delay starts. When the delay time
expires, the Alarm System is rearmed; if the temperature and/or CO

2

should fall below

or rise above the alarm setpoints, the alarm will be activated. If chamber conditions
recover within the Alarm Arming Delay time, the Alarm System will be rearmed but no
alarm will be activated.

If an alarm is not acknowledged but the chamber conditions subsequently recover, the
audible alarm will be cancelled but the alarm message will remain on the screen to alert
the user to the fact that an alarm has occurred. You can cancel this alarm message by
pressing any key. The duration of the audible alarm can be adjusted from inactive to
continuous (see Section 3.7.4 above).