Eppendorf S Series 48S CO2 Incubator User Manual
Page 10

10
CO48S-0050
Galaxy S Series 48 S Personal Size CO
2
Incubator
User’s Guide
The Galaxy 48 S model contains many standard features usually seen as options. In addition to
its viewing window to allow visualization of the cultures without compromising the internal
atmosphere, there is a 25mm access port now standard to allow for seamless integration of
independent probes or other equipment through the chamber. NBS has also included an RS232
port as standard on all Galaxy S series incubators. This port will communicate with any
computer through a hyperlink access or can be used to externally datalog the incubator through
NBS software.
The incubator incorporates a two-level alarm system. The chamber-monitoring alarms are
programmable and will alert you if temperature or CO
2
have not recovered within a preset time
after the door has been opened. If it is not required, this system can be disarmed. The system
alarms occur only if a system component problem has developed that requires user intervention
to rectify. The incubator also incorporates an over-temperature safety system that operates
independently from the main control system.