New Brunswick Scientific

User’s Guide

3. Press the

enter key to save the desired value.



3.7.4 Alarm

Duration

The following steps will adjust the length of time during which all audible
alarms
will last:

1. In the alarm menu (see step 1 of Section 3.7.1 above), press the S

key. The

display shows

Per

.

Iod

(reading across both displays).

2. Press the

enter key and the S & T keys to adjust the Alarm duration (as

you scroll through the available choices, you will see

OFF

,

10 sec

,

30 sec

,

60 sec

,

600 sec

,

1 Hr

, then

ON

).

3. Press the

enter key to select the desired value.


3.7.5

Alarm Arming Delay

This is the length of time that is allowed for the Temperature and CO

2

to recover

after you open the incubator, and before the Alarm System is armed again. This
helps prevent unnecessary alarms from occurring. To change the delay:

1. In alarm menu (see step 1 of Section 3.7.1 above), press the S key. The

display will show

dr

.

dEL

(across both windows).

2. Press the

enter key and the S & T keys to adjust the Alarm Arming time

(as you scroll through the available choices, you will see

0

.

15 Hr

,

0

.

20

,

0

.

30

,

1

.

00

, then

OFF

).

3. Press the

enter key to select the desired value.


Press ¤ to return to the main alarm programming menu, and press ¤ again to
return to the main display.

NOTE:

When alarm delay is set to OFF, the Alarms will arm only when the
programmed setpoint is reached.


3.8 Chamber Alarm System

When the incubator is switched

ON

, or after any values have been reprogrammed, the

Alarm System is inactive until the setpoint values (± 0.1) are achieved, after which the
Alarm System is armed. See Figure 7, Chamber Alarm System Flow Chart, on the
following page.