beautypg.com

3o o – Eppendorf S Series 48S CO2 Incubator User Manual

Page 17

background image

17

New Brunswick Scientific

User’s Guide

3

3

O

O

P

P

E

E

R

R

A

A

T

T

I

I

O

O

N

N

3.1 Control

Panel

The control panel consists of two LCD displays and four function keys (see Figure 5):

Figure 5: Control Panel

1 Temperature

display

2

CO

2

display

3

Function keys

Table 3: Function Keys

Key Symbol

Key Name

Function

¤

Programming

Press this key to enter Programming mode, and to
set values in either display. Press this key
simultaneously with the Enter key to perform
AutoZero.

T

Down

In Programming mode, use this key to scroll down
through numbered values in the display. Press this
key simultaneously with the Enter key to work in
the Alarm system.

S

Up

In Programming mode, use this key to scroll up
through numbered values in the display.

enter

Enter

Press this key to save a new setpoint. Press this
key simultaneously with the Programming key to
perform AutoZero. Press this key simultaneously
with the Down key to work in the Alarm system.

NOTE:

If you accidentally press both the Up and Down keys simultaneously, you
will engage Engineering Mode: immediately press the Programming key
(

¤) to exit.

2

1

3