Excella
®
E-25/E25R Classic Incubator Shakers
Operating Manual
Symptom(s)
Probable Cause(s) & Solution(s)
Load out of balance: unload all contents, then reload.
Operating noise
Loose component(s) in platform, subplatform and/or drive
assembly: call for service.
Heater fuse blown: replace.
Compressor fuse blown: replace.
Compressor over-pressure switch activated: call for service.
Ambient temperature too high or too low: cool or heat the room
as needed.
Defective heater: call for service.
Defective refrigeration system: call for service.
Incubator does not
reach set temperature.
Incorrect temperature indication (see below).
Defective RTD assembly: call for service.
Incorrect temperature
indication.
Defective main board: call for service.