27

New Brunswick Scientific

Operating Manual

6.10 Maximum Recommended Speed

Figure 6 shows the maximum recommended speed for your shaker, according to load.

Figure 6: Load & Speed for E-25/E-25R

0

50

100

150

200

250

300

350

400

<30

30 to 36

36 to 40

40 to 60

>60

Platform and Load (lb)

50,

125,

250 mL

20% Full

1L,

2L,

2.8L

20% Full

NOTE:

In Figure 6, “20% Full” refers to the amount of liquid in the flasks. The
platforms are fully loaded with flasks.