New Brunswick Scientific

Operating Manual

4. Set the speed by using the 

(UP)

or

(DOWN) KEY

until the desired setpoint is

displayed. Holding the 

(UP)

or

(DOWN) KEY

will cause the setting to change

more rapidly.

NOTE:

The setpoint may be changed during a run without stopping the shaker by
following steps 2 - 4 above. During speed changes, a visual alarm (flashing

RPM INDICATOR

) will flash and an audible alarm will sound until the speed

returns to within 5 rpm of the setpoint.

6.3 Checking a Setpoint

1. Press

SELECT

until the desired indicator is illuminated.

2. Press either 

(UP)

or

(DOWN) KEY

to enter the

SET MODE

and display the current

setpoint.

CAUTION!

Holding the

(UP)

or

(DOWN) KEY

for more than 0.5 seconds causes the

speed setpoint to change. Should this occur, resetting will be necessary.

6.4 Timed

Functions

The shaker may be programmed to automatically stop after a preset time period of 0.1 to
99.9 hours. There must be power to the shaker in order to set the timer, although a timed
run can be initiated while the shaker is either stopped or operating.

To set the timer:
1. Press the

SELECT KEY

until the

HRS INDICATOR

is illuminated.

2. Press 

(UP)

or

(DOWN) KEY

to enter the

SET MODE

and set the desired run time,

between 0.1 and 99.9 hours.

If the shaker is stopped, skip to Step 5 below. If the shaker is already running:

3. Press the

START/STOP KEY

. The shaker will stop and the display will read

OFF

.

4. Press the

START/STOP KEY

again; the

TIME INDICATOR

will light and the shaker will

start the timed run.