9 Technical data

28

ThermoStat plus — Operating Manual

9.4 Application parameters

9.5 Additional data

9.5.1

Interfaces

9.5.2

Time interval

Temperature control range:

–5 °C to 99 °C

Usable tempering range:

from 30 °C below room temperature to 99 °C

Usable tempering range CombiBox:

30 °C below room temperature to 95 °C
(dependent on filling level)

Tempering accuracy for:

0 °C

±1.0 °C

37 °C

±0.5 °C

90 °C

1.0 °C

Heating rate:

5 °C/min

Cooling rate:

simple tempering operation:

6.0 °C/min between 99 °C and 25 °C

1.5 °C/min between 25 °C and –5 °C

programmable tempering operation:

1 to 2 °C/min

Temperature deviation across the 24

exchangeable thermoblock:

max. 0.2 °C

EDP connection:

RS-232, Sub-D9 male

Hint!

Only connect devices to interfaces that meet the IEC 950/EN 60950 (UL 1950)

standards.

Programmable time interval:

1 min to 99:59 hours, infinitely adjustable