5 Operation
22
ThermoStat plus — Operating Manual
4. Set the desired tempering duration for this phase using the "Time" arrow keys.
5. Press the "Prog." key.
On the display the parameters for Phase 2 appear.
6. Set the parameters as desired using the arrow keys.
7. Repeat steps 5 and 6 to set the parameters for Phase 3 and Phase 4.
The timer in the programmable mode works continuously, i.e. the device does not take into
account the time required to reach the next temperature. You can program these so-called
temperature change phases if you have set the parameters for the phases:
8. Press the "Prog." key to change to the temperature change phases between the program
phases.
On the display the first temperature change phase T1-2 appears on the left, to its right the
target temperature of phase 2 and the current actual temperature. In the bottom line the
tempering duration of this phase is shown. The duration is always automatically disabled
("--h: --m").
9. Enable the temperature change phase using the "Time" arrow keys.
The minimum time the device needs to change from the temperature of phase 1 to the
temperature of phase 2 is shown.
10. Press the "Time" arrow keys to extend the temperature change phase or disable it, if
necessary.
11. Press the "Prog." key.
On the display the parameters for the second temperature change phase T2-3 are shown,
which you can modify or enable as described above.
12. Press the "Prog." key.
On the display the parameters for the third temperature change phase T3-4 are now shown,
which you can modify or enable as described above.
13. Press the "Prog." key.
The programming is complete and the parameters of the first phase (P1) are shown.
14. Press the "Start/Stop" key to start the programmed tempering.
The program step shown in the display will always be started. The phases then run
automatically sequenced until phase 4 has been completed. A maximum of 4 phases can be
sequenced automatically, however, without roll-over (not via P4 back to P1). The values are
retained even after the device has been switched off.