7 Maintenance

25

ThermoStat plus — Operating Manual

7.2 Disinfection / Decontamination

Auxiliary equipment

lint-free cloth,

disinfectant.

1. Switch the ThermoStat plus off and isolate from the power supply.
2. Allow the device to cool down.
3. Clean the device (see Cleaning on page 24).
4. Select a disinfection method which complies with the legal requirements and regulations

applicable to your range of application.

5. Wipe the surfaces with the lint-free cloth and disinfectant.

7.3 Decontamination before shipping

If you are shipping the device to the authorized Technical Service for repairs or to your authorized

dealer for disposal please note the following:

7.4 Temperature control

The application parameters in the technical data (see p. 28) define the limits and deviations

(tolerances) on all positions of a thermoblock / exchangeable thermoblock.
Use the temperature validation system for an exact control of the temperatures. For the 1.5 ml

reference exchangeable thermoblock there is a temperature sensor - 1.5 mL for all Thermomixer

and ThermoState from Eppendorf which can measure the exact temperature in the block

(see Temperature sensor on page 31).

DANGER!

Danger! Electric shock as a result of penetration of liquid.

Switch off the device and disconnect it from the power supply before starting

cleaning or disinfecting.

Do not allow any liquids to enter the inside of the housing.

Do not perform spay disinfection.

Only reconnect the device to the power supply once it is completely dry.

WARNING!

Warning! Risk to health from contaminated device

1. Follow the instructions in the decontamination certificate. It is available in PDF

format on our homepage (www.eppendorf.com/decontamination).

2. Decontaminate all the parts you want to dispatch.
3. Enclose the fully-completed decontamination certificate for returned goods (incl. the

serial number of the device) with the dispatch.