3 safety – Eppendorf ThermoStat plus User Manual
Page 11
3 Safety
13
ThermoStat plus — Operating Manual
WARNING!
Warning! Contamination due to opening tube lids.
In the following cases, the lids of PCR tubes can spring open and the sample material
can escape:
•
high vapor pressure of the content
•
improperly sealed lid
•
damaged sealing lip
Always check that micro test tubes have been sealed tightly before use.
WARNING!
Warning! Risk to health from contaminated device
Perform decontamination before storing or dispatching the device and/or its
accessories.
CAUTION!
Caution! Device damage from improper use.
Only use the ThermoStat plus with the exchangeable thermoblock.
CAUTION!
Caution! Damage to the display due to mechanical pressure.
Do not exert mechanical pressure on the display.
CAUTION!
Caution! Damage to electronic components from condensation.
After moving the device from a cooler environment (e.g: cool room or outdoors),
wait at least an hour before connecting it to the mains power supply.
CAUTION!
Caution when using aggressive chemicals.
Aggressive chemicals may damage both the device and its accessories.
Do not use any aggressive chemicals on the device and accessories such as strong
or weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or
phenol.
If the device becomes contaminated with aggressive chemicals, clean it
immediately with a neutral cleaning agent.
CAUTION!
Caution! Lack of safety due to missing operating manual.
When passing on the device, always enclose the operating manual.
If you lose the operating manual, request a replacement. The current version of the
operating manual and the safety instructions can also be found on our website
www.eppendorf.com.