9 technical data, 1 ambient conditions, 2 errors – Eppendorf Top Buret M / H User Manual
Page 36: 3 parameters for titratable media, 4 conditions for calibration, Technical data 9.1, Ambient conditions, Errors, Parameters for titratable media, Conditions for calibration

36
Technical data
Top Buret
™
English (EN)
9
Technical data
9.1
Ambient conditions
9.2
Errors
9.3
Parameters for titratable media
9.4
Conditions for calibration
Test conditions and test evaluation in accordance with ISO 8655, part 6. Test
with analytical balance with evaporation protection, approved by the board of
weights and measures.
The errors were determined under the following conditions:
Ambience
Only for use in indoors.
Ambient temperature
15 °C – 40°C
Relative humidity
0 % – 90 %, non-condensing.
Atmospheric pressure
Use up to a height of 2000 m
above sea level.
Model
Testing volume
Inaccuracy
Imprecision
Top Buret M
25 mL
±0.2%
≤ 0.1 %
Top Buret H
50 mL
±0.2%
≤ 0.1 %
Medium temperature
15 °C – 40°C
Maximum density
2.2 g/cm
3
Maximum vapor pressure
50 kPa
Liquid
Water according to ISO 3696
Number of determinations
10
Ambient temperature
20 °C – 25°C ± 0.5 °C
Dispensing liquid
Dispensing on the tube inner wall