beautypg.com

9 technical data, 1 ambient conditions, 2 errors – Eppendorf Top Buret M / H User Manual

Page 36: 3 parameters for titratable media, 4 conditions for calibration, Technical data 9.1, Ambient conditions, Errors, Parameters for titratable media, Conditions for calibration

background image

36

Technical data
Top Buret

English (EN)

9

Technical data

9.1

Ambient conditions

9.2

Errors

9.3

Parameters for titratable media

9.4

Conditions for calibration

Test conditions and test evaluation in accordance with ISO 8655, part 6. Test
with analytical balance with evaporation protection, approved by the board of
weights and measures.

The errors were determined under the following conditions:

Ambience

Only for use in indoors.

Ambient temperature

15 °C – 40°C

Relative humidity

0 % – 90 %, non-condensing.

Atmospheric pressure

Use up to a height of 2000 m 
above sea level.

Model

Testing volume

Inaccuracy

Imprecision

Top Buret M

25 mL

±0.2%

≤ 0.1 %

Top Buret H

50 mL

±0.2%

≤ 0.1 %

Medium temperature

15 °C – 40°C

Maximum density

2.2 g/cm

3

Maximum vapor pressure

50 kPa

Liquid

Water according to ISO 3696

Number of determinations

10

Ambient temperature

20 °C – 25°C ± 0.5 °C

Dispensing liquid

Dispensing on the tube inner wall