Page 26: Emptying the device, 5 removing the bottle 5.6 emptying the device

background image

26

Operation
Top Buret

English (EN)

5.5

Removing the bottle

5.6

Emptying the device

Before changing the liquid and before cleaning, the Top Buret must be emptied.

Proceed as follows:

Prerequisites
• The bottle has been unscrewed (see p. 26)

1. Mount the Top Buret onto an empty bottle.
2. Place a collection vessel below the canula.
3. Rotate the discharge valve toggle to titration.
4. Rotate the dispensing wheels forward until no more liquid runs out of the

discharge tube.

5. Rotate the discharge valve toggle to deairing.
6. Rotate the dispensing wheels forward until no more liquid runs out of the

recirculation tube.

Abb. 5-4: Emptying the telescopic aspirating tube

Fig. 5-4:

Emptying the telescopic
aspirating tube

1. Rotate the discharge valve toggle to

deairing.

2. Place the Top Buret with bottle into

a suitable collecting vessel.

3. Unscrew the device from the bottle

in a counterclockwise direction.

4. Lift the Top Buret until the

telescopic aspirating tube is no
longer immersed in the liquid.

5. Gently tap the telescopic aspirating

tube against the inside of the bottle.
The residual liquid from the
telescopic aspirating tube runs into
the bottle.

To remove the reagents completely, clean the device.