7 troubleshooting, 1 troubleshooting, Troubleshooting 7.1 – Eppendorf Top Buret M / H User Manual

Page 31: Troubleshooting, 7troubleshooting 7.1 troubleshooting

background image

31

Troubleshooting

Top Buret

English (EN)

7

Troubleshooting

7.1

Troubleshooting

Symptom/message

Cause

Remedy

• Air is aspirated.
• The titrated liquid

contains air bubbles.

• The device was not

sufficiently deaired.

Deair the device 
(see p. 23).

• The telescopic

aspirating tube has
not been mounted
completely.

Slide the telescopic
aspirating tube firmly onto
the connection opening 
(see p. 18).

• The telescopic

aspirating tube is
damaged.

Shorten the telescopic
aspirating tube or replace
it.(see p. 18).

• The telescopic

aspirating tube does
not immerse into the
liquid.

Lengthen the telescopic
aspirating tube until it is
immersed in the liquid 
(see p. 18).

• The connection

opening is damaged.

Send in the device for repair.

Liquid is not aspirated. • The connection

opening is clogged.

Clean the device 
(see p. 32).

The titration volume is
too small.

• The telescopic

aspirating tube has
not been mounted
completely.

Slide the telescopic
aspirating tube firmly onto
the connection opening (see
p. 18).

• The telescopic

aspirating tube is
damaged.

Shorten the telescopic
aspirating tube or replace it
(see p. 18).

• The device is

incorrectly adjusted.

Set the device to the factory
adjustment (see p. 30).

• The connection

opening is damaged.

Send in the device for repair.

The display shows the

symbol.

• The batteries are flat.

Replace the batteries.