Deairing the Top Buret
Air bubbles in the supply lines and in the feeding mechanics will distort the
measuring results.
Deair the Top Buret in the following situations:
• Prior to initial operation
• After intensive cleaning
• After changing the bottle
• After a long period of inactivity
Proceed as follows:
Prerequisites
• The device has been completely mounted.
CAUTION! Personal injury due to contact with reagents.
Contact with reagents may be harmful to eyes or skin.
Wear personal protective equipment.
Do not point the discharge tube at people.
Abb. 5-1: Deairing the interior system
Fig. 5-1:
Deairing the interior system
1. Rotate the discharge valve toggle to
deairing.
2. Remove the control panel.
3. Rotate the dispensing wheels slowly
and evenly forward until there is no
more air left in the system.
- epMotion 96 (76 pages)
- epMotion 5070 (100 pages)
- epMotion 5075 (130 pages)
- Centrifuge 5427 R (64 pages)
- Centrifuge 5427 R (104 pages)
- White Paper 14 (8 pages)
- Rolling Cabinet (34 pages)
- Mastercycler nexus (142 pages)
- Mastercycler nexus (118 pages)
- Concentrator plus (New Design) (48 pages)
- Concentrator plus (43 pages)
- Easypet 3 (38 pages)
- Xplorer (74 pages)
- Xplorer Adjustment (26 pages)
- AF2200 Plate Reader (72 pages)
- AF2200 Plate Reader (78 pages)
- G0.5 µPlate (32 pages)
- BioSpectrometer basic (104 pages)
- BioSpectrometer kinetic (106 pages)
- BioSpectrometer fluorescence (102 pages)
- Micro Test Tubes (5 pages)
- Microplates (10 pages)
- PiezoXpert (34 pages)
- Eporator (38 pages)
- MiniSpin (20 pages)
- MiniSpin (25 pages)
- Centrifuge 5702 (32 pages)
- 5702 Centrifuge (27 pages)
- 5702 Centrifuge (32 pages)
- C5702 RH Centrifuge (32 pages)
- 5418 Centrifuge (80 pages)
- 5418 Centrifuge (48 pages)
- 5424 Centrifuge (44 pages)
- 5424 Centrifuge (71 pages)
- 5430 Centrifuge (88 pages)
- 5430 Centrifuge (130 pages)
- 5804 Centrifuge (95 pages)
- 5804 Centrifuge (127 pages)
- 5804 Centrifuge (129 pages)
- TransferMan4 r (102 pages)
- TransferMan4 m (96 pages)
- InjectMan 4 (100 pages)
- InjectMan NI 2 (60 pages)
- InjectMan NI 2 (16 pages)
- PatchMan NP 2 (53 pages)