3 application parameters, 1 celltram air, 2 celltram oil – Eppendorf CellTram vario User Manual
Page 26: 3 celltram vario, Celltram® air · oil · vario — operating manual
CellTram® Air · Oil · vario — Operating manual
26
7.3 Application parameters
7.3.1
CellTram Air
7.3.2
CellTram Oil
7.3.3
CellTram vario
Change in volume
(per rotary knob revolution):
88 μL
Minimum adjustment volume
< 0.2 μL
Adjustable total volume
2.640 μL
Entire stroke of the piston
15 mm (corresponds to 30 revolutions)
Generation of pressure
Using piston and cylinder
Change in volume
(per rotary knob revolution):
9,6 μL
Minimum adjustment volume:
< 0.02 μL
Adjustable total volume
960 μL
Total filling volume
(incl. 1 m tubing and universal
capillary holder)
< 2 mL
Entire stroke of the piston
50 mm (corresponds to 100 revolutions)
Generation of pressure
Using piston and cylinder
Change in volume (per rotary knob
revolution)
Coarse: 9.6 μL
Fine: 0.96 μL
Minimum adjustment volume
Coarse: < 0.02 μL
Fine: < 0.002 μL
Adjustable total volume
960 μL
Total filling volume
(incl. 1 m tubing and universal
capillary holder)
<2 mL
Entire stroke of the piston
50 mm (corresponds to 100 revolutions)
Generation of pressure
Using piston and cylinder