Celltram® air · oil · vario — operating manual – Eppendorf CellTram vario User Manual
Page 23

CellTram® Air · Oil · vario — Operating manual
23
Required equipment
•
Alcohol (ethanol, isopropanol) or disinfectants containing alcohol
Proceed as follows:
1. Choose the disinfection method which corresponds to the legal regulations and guidelines in
place for your range of application.
2. Wipe all surfaces of the CellTram and accessories using the disinfectant.
3. Flush the pressure tube and universal capillary holder with the disinfectant if required.
4. To remove residual liquid, blow out the pressure tube and universal capillary holder with
compressed air after cleaning.
Hint!
There must not be any residual liquid remaining in the pressure tube or in the universal
capillary holder after cleaning as otherwise steady pressure behavior will be
compromised.