CellTram® Air · Oil · vario — Operating manual

15

4.3 Filling oil

Valid for:

CellTram Oil

CellTram vario

Abb. 9:CellTram Oil and CellTram vario

Fig.

9: CellTram

Oil and CellTram vario

Hint!

Work with an air cushion in the visible area of the capillary in order to avoid contact

between the oil and the medium.

Hint!

Observe the legal regulations for purity of the oil. The supplied oil is generally used in

the ICSI area to coat drops containing embryos / egg cells or sperm. The supplied oil is

not sterile. Additional instructions can be found in the safety data sheet of the

manufacturer.

Hint!

Aspirating oil with the syringe may lead to the formation of tiny air bubbles. This should

be avoided as these air bubbles may be transferred to the CellTram, which may impair

or slow down the filling process. If the syringe contains air bubbles, these should be

collected by long-term storage. Large air bubbles can be pressed out during filling.

1 Initial position of piston

2 Filling bore with plug screw

3 Final position of piston

4 Barb screw

5 Pressure tubing connection to the rear

knurled screw

6 Grip head 0 for capillary

1 2

3

4

5

6