beautypg.com

1 preparing the external pressure supply, 2 connecting the compressor, Preparing the external pressure supply – Eppendorf FemtoJet 4x v.2 User Manual

Page 21: Connecting the compressor

background image

Installation

FemtoJet

®

4x

English (EN)

21

4.3.1

Preparing the external pressure supply

For connecting the Microinjector, mount the appropriate quick coupling to the pressure
supply. The installation only must be realized by qualified staff who has received training
on compressed air systems.

Prerequisites

• Quick coupling from brass with conical thread is available (delivery package).

• Nickel plated quick coupling with cylindrical thread is available (delivery package).

Install the quick coupling.

4.3.2

Connecting the compressor

Prerequisites

• Quick coupling (nickel plated) is mounted to the compressor.

• The Microinjector is switched off.

1. Connect the pressure tube to the compressor.

2. Connect the pressure tube to the Microinjector.

3. Switch on the compressor.

CAUTION! Material damage due to contaminated compressed air or
compressed gas
Compressed air containing oil or contaminations will damage the valves and
sealings in the Microinjector.

Only use oil-free, clean and dry compressed air.

Only use cleaned compressed gas (e.g. nitrogen gas).