System management

Mastercycler

®

nexus

English (EN)

79

 Press the

OK

softkey.

The editor is displayed.

 Press the

>>

softkey. Press the

Save

softkey.

You can start the program without renewed optimization of the parameters.

8.7

Network connection

You can connect the Mastercycler nexus to an Ethernet network to send e-mails and log files. To set up the
network connection, you will need the following information from your network administrator:

• IP address, subnet mask and gateway address for the Mastercycler nexus (specified by the network

administrator).
Alternatively, you can assign the IP address automatically via DHCP.

• IP address of the primary DNS server

You can optionally enter the IP address of a backup DNS server. Alternatively, you can assign the IP
address automatically via DHCP.

• For sending e-mails:

– E-mail address of the sender with the user name and password to log into the SMTP server (if

required)

– Name or address and port number of the SMTP server

• For sending Syslog messages:

– Name or address and port number of the syslog server

Setting from the Mastercycler 533x program

Corresponding setting in the program header

LID 105°C

Lidtemp.

WAIT

TSP heated lid

AUTO

switch off lid at low blocktemperature

CNTRL

Block

=

Temp. mode

fast

Tube

=

Temp. mode

standard

or

safe

Select the

Temp. mode

setting based on the reaction

volume or the sample properties (see p. 52). The
temperature control mode for 0.5 mL standard
tubes (not thin-walled) is no longer supported on
the Mastercycler nexus.

Block temperature control behavior

Check the

Simulate Mastercycler gradient

checkbox.