1 cycler system functions, Cycler system functions – Eppendorf Mastercycler nexus User Manual
Page 74
System management
Mastercycler
®
nexus
English (EN)
74
8.4.1
Cycler system functions
The
System
node contains the cycler system level with the following softkeys.
Status
• Display the cycler operating status
• Display the temperatures of the thermoblock and the heated lid
• Display the remaining run time for the program
• Stop the program
Protocol
Protocols of the last 20 program runs:
Select the protocol via the arrow keys.
•
softkey: print the protocol
•
Export
softkey: copy the protocol to a storage medium
•
Show
softkey: display the protocol
The following applies to devices with a thermoblock with 64+32 wells:
The Cycle functions comprise the
Properties
,
Overview
,
Status 64
and
Status 32
softkeys.
The
Incubate
and
Protocols
softkeys are located under the subnodes
Block 64
or
Block 32
.
Info
• Display the device name
• Display the date of the last verification
• Display the date of the last verification
• Display the programs run (since the last verification)
• Display the date of the last self-test
• Display the date of the next self-test
• Display the programs run (since the last self-test)
• Display the date of the last adjustment
Details about the next verification or self-test are only given if the settings have been
made under
Service intervals
Restart
Entry for maximum duration of interruption of voltage supply (
mm:ss
) after which a
running program should be automatically restarted. The program is not restarted after
a longer interruption.
Sensors
Display current sensor temperatures for the control loops of the thermoblock and the
heated lid.