System management
Mastercycler

®

nexus

English (EN)

74

8.4.1

Cycler system functions

The

System

node contains the cycler system level with the following softkeys.

Status

• Display the cycler operating status
• Display the temperatures of the thermoblock and the heated lid
• Display the remaining run time for the program
• Stop the program

Protocol

Protocols of the last 20 program runs:

 Select the protocol via the arrow keys.

Print

softkey: print the protocol

Export

softkey: copy the protocol to a storage medium

Show

softkey: display the protocol

The following applies to devices with a thermoblock with 64+32 wells:
The Cycle functions comprise the

Properties

,

Overview

,

Status 64

and

Status 32

softkeys.

The

Incubate

and

Protocols

softkeys are located under the subnodes

Block 64

or

Block 32

.

Info

• Display the device name
• Display the date of the last verification
• Display the date of the last verification
• Display the programs run (since the last verification)
• Display the date of the last self-test
• Display the date of the next self-test
• Display the programs run (since the last self-test)
• Display the date of the last adjustment

Details about the next verification or self-test are only given if the settings have been
made under

Service intervals

((see p. 70)).

Restart

Entry for maximum duration of interruption of voltage supply (

mm:ss

) after which a

running program should be automatically restarted. The program is not restarted after
a longer interruption.

Sensors

Display current sensor temperatures for the control loops of the thermoblock and the
heated lid.