34

- Wear protective gloves during cleaning the instrument.
- Electrical devices may not be placed in the cleaning agent

for the purpose of cleaning.

- Do not allow moisture to get into the equipment when

cleaning.

- Please consult IKA

®

before using any cleaning or decon-

tamination methods, other than those recommended here.

- The handles can not be place in the cleaning agent for

cleaning.

- The only cleaners or disinfectants that may be used are

those that:

- lie in the pH range 5 - 8,
- contain no corrosive alkalis, peroxides, chlorine com-

pounds, acids or brine.

Fig. 22

Only clean IKA

®

appliances using these IKA

®

approved

cleaning agents.

Dirt

Cleaning agent

Dyes

Isopropanol

Building materials Water containing detergent, Isopropanol
Cosmetics

Water containing detergent, Isopropanol

Food

Water containing detergent

Fuels

Water containing detergent

Other materials

Please consult IKA

®

Spare parts order

When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
- item and designation of the spare part see www.ika.

com, spare parts diagram and spare parts list.

Repair

Please send equipment for repair only after it has

been cleaned and is free from any materials which

may constitute a health hazard.

For repair, please fill up the “Decontamination Certifi-

cate” form supplied with instrument or printed copy on the

IKA

®

website: www.ika.com.

If you require servicing, return the equipment in its original
packaging. Storage packaging is not sufficient. Please also
use suitable transport packaging.

Error codes

The fault is shown by an error message in the display as following if the error occurs.

Proceed as follows in such cases:

Switch off the device using the main switch

Carry out corrective measures

Restart the instrument

Error code Cause

Effect

Corrective action

Error 3

• Permitted ambient temperature has

been exceeded

• Ventilation slots or fan housing is

blocked

Heating off • Switch off the unit. Allow it to cool down and

then switch on again.

• Clean ventilation slots or fan housing
• Observe maximum permissible ambient

temperature

Error 4

• Motor is blocked or overloaded

Motor off

• Reduce the torque load (load)
• Reduce the setpoint speed

Error 8

• Fault in calibration procedure
• Value was incorrectly stored to

memory

Heating off • Repeat the calibration procedure

Error 14

• Short-circuit at temperature sensor

plug

• Short-circuit in connecting cable or on

temperature sensor

Heating off • Check the plug

• Substitude the temperature sensor