Maintenance and cleaning – IKA LR 1000 control Package User Manual
Page 33
33
Maintenance and cleaning
Cleaning
For cleaning disconnect the main plug!
Disassemble the reactor vessel:
Remove the reactor vessel from the LR control base.
Remove the reactor cover from the reactor vessel.
Remove the glass tube from the base as described below.
Use a screwdriver to loosen and remove the four screws on
the two clamps.
Then, both handles can be removed from the glass vessel.
Note: the screws can only be loosened and can not be re-
moved from the handles.
Now both clamps can be removed. Then remove the glass
tube with handles from the base.
Remove both handles from the glass vessel as following.
Reserve the glass tube with handles. Loosen both screws
with screw driver at the bottom of handles (see Fig. 19).
Disassemble the anchor stirrer:
After removing the glass tube from the vessel base, the an-
chor stirrer together with the reactor vessel bottom of the
socket can be taken apart from the vessel base (see Fig. 20).
Please pay attention to the O-ring when you disassemble
the reactor vessel.
Open the screw as described in Fig. 21 with the socket
wrench.
Now you can remove the anchor stirrer as shown in Fig. 22.
All O-rings can now be carefully removed by hand or using a
blunt tool for cleaning too.
Fig. 18
Fig. 19
O-ring
Fig. 20
Fig. 21