beautypg.com

Class1 Intelli Tank level display with drip empty User Manual

Page 10

background image

Intelli-Tank Manual (106759)092204

– 10 –

Class 1

2 Point Calibration

1. Enter the password RLLR LLRL. The unit will respond by flashing the two

center LEDs twice. The unit will then begin cascading the LEDs from top to
bottom (drip).

2. Make certain that the tank is EMPTY and then activate the RIGHT switch to

store that point. The unit will flash the top LED and then turn on all four LEDs.

3. Fill the tank and then activate the RIGHT switch. The unit will respond by

flashing the top LED then lighting the two center LEDs and then reverting to
normal operation by displaying FULL (all LEDs on).

5 Point Calibration

1. Enter the password RLLR LRLR. The unit will respond by flashing the two

center LEDs five times. The unit will then begin cascading the LEDs from
top to bottom (drip).

2. Make certain that the tank is EMPTY and then activate the RIGHT switch to

store that point. The unit will flash the top LED and then turn on the bottom
LED.

3. Fill the tank to the one-quarter tank point and then activate the RIGHT switch.

The unit will flash the top LED and then turn on the bottom two LEDs.

4. Fill the tank to the one-half tank point and then activate the RIGHT switch.

The unit will flash the top LED and then turn on the bottom three LEDs.

5. Fill the tank to the three-quarter tank point and then activate the RIGHT

switch. The unit will flash the top LED and then turn on all four LEDs.

6. Fill the tank to the full point and then activate the RIGHT switch. The unit will

respond by flashing the top LED then lighting the two center LEDs and then
reverting to normal operation by displaying FULL (all LEDs on).

9 Point Calibration

1. Enter the password RLLR RLLR. The unit will respond by flashing the two

center LEDs nine times. The unit will then begin cascading the LEDs from
top to bottom (drip).

2. Make certain that the tank is EMPTY and then activate the RIGHT switch to

store that point. The unit will flash the top LED and then begin flashing the
bottom LED.

3. Fill the tank to the one-eighth point and then activate the RIGHT switch. The

unit will flash the top LED and then turn on the bottom LED.

4. Fill the tank to the one-quarter tank point and then activate the RIGHT switch.

The unit will flash the top LED and then turn on the bottom LED and flash the
second LED.

5. Fill the tank to the three-eighths point and then activate the RIGHT switch.

The unit will flash the top LED and then turn on the bottom two LEDs.