beautypg.com

Braun Sterile Container System 9-Jul User Manual

Page 9

background image

7

Only process chemicals that have been tested and approved (e.g. VAH/
DGHM or FDA approval or CE mark) and which are compatible with the
product’s materials according to the chemical manufacturers’
recommendations may be used for processing the product. All process
parameters specified by the chemical’s manufacturer, such as
temperatures, concentrations and exposure times, must be strictly
observed. Failure to do so can result in the following problems:

• optical changes of materials, e.g. fading or discoloration of aluminum.

For aluminum, pH >8 in the application/process solution is sufficient
to cause visible surface changes.

• material damage such as corrosion, cracks, fracturing, premature aging

or swelling.

¾ Do not use process chemicals that cause stress cracking of plastics

such as PPSU, or attack softeners e.g. in silicone and cause brittleness
of the material.

5.2 Cleaning/Disinfecting

CAUTION

Damage to the product due to inappropriate
cleaning/disinfecting agents and/or excessive
temperatures!
¾ Use cleaning and disinfecting agents according

to the manufacturer’s instructions. The cleaning
and disinfecting agent must
- be approved for plastics and high-grade
steel
- not attack softeners (e.g. in silicone).

¾ Observe specifications regarding concentration,

temperature and exposure time.