Sterile container system

Aesculap Sterile Technology

5.3 Manual cleaning/disinfecting

¾ Inspect visible surfaces for residual contamination after manual

cleaning/disinfecting.

¾ Repeat the cleaning process if necessary.

Manual cleaning and wipe disinfection

D–W:

Drinking water

FD–W:

Fully desalinated water (demineralized)

RT:

Room Temperature

Stage I

¾ Wipe off gross debris prior to cleaning.

Stage II

¾ Dry the product with lint-free tissue or medical-quality filtered

compressed air.

Stage III

¾ Carry out wipe disinfection.

Stage IV

¾ When using aldehyde-free disinfectants:

After the specified exposure time (at least 1 minute), rinse the
disinfected surfaces under running FD water.

¾ Allow water to drip off for a sufficient length of time.

Stage V

¾ Dry the product with lint-free tissue or in a heating cabinet.

Stage

Step

T

[°C/°F]

t

[min]

Conc.

[%]

Water

quality

Chemical

I

Cleaning

RT

(cold)

-

-

D–W

-

II

Drying

RT

-

-

-

-

III

Wipe disinfection

-

>

1

-

-

I
Alcohol Denat. 70 %
(ethanol B|BRAUN)

II
Aldehyde-free surface
disinfectant (Stabimed)

IV

Final rinse

RT

(cold)

-

-

FD–W

not required

Rinse off cleaning
chemicals; do not leave
residues

V

Drying

RT

-

-

-

-

-