Switch on continuous transmission in current unit, Switch off continuous transmission in current unit, Send all implemented commands – RADWAG Scales of WPT series User Manual
Page 71

- 71 -
Possible answers:
C0_I CR LF
- command comprehended but cannot be executed
C0_A CR LF
- command comprehended and executed
19.4.10. Switch on continuous transmission in current unit
Syntax: CU1 CR LF
Possible answers:
CU1_I CR LF
- command comprehended but cannot be executed
CU1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in current unit is returned
Frame format:
1 2 3
4
5
6 7-15 16 17 18 19 20 21
S U I stability space sign mass space
unit
CR LF
19.4.11. Switch off continuous transmission in current unit
Syntax: CU0 CR LF
Possible answers:
CU0_I CR LF
- command comprehended but cannot be executed
CU0_A CR LF
- command comprehended and executed
19.4.12. Send all implemented commands
Syntax: PC CR LF
Possible answers:
PC_- >_Z,T,TO,S,SI,SU,SUI,C1,C0,CU1,CU0,PC – command executed,
the indicator have sent all the implemented commands.