Zeroing, Weighings in two ranges, Eighings in two ranges – RADWAG Scales of WPT series User Manual
Page 26

• Using
and
set the tare value,
• Press ,
• Program returns to weighings mode. The inscribed tare value
can be seen on the display with „–” sign,
• Tare can be inscribed anytime in weighings mode.
Notice:
1. You cannot inscribe a new tare value when the tare value in memory
is greater than zero. In the case of trying this the
will be displayed and short audible signal will be emitted.
2. Users can also enter up to 9 tare values to the scale memory
(see 15.10 of his manual).
11.3. Zeroing
To ZERO the scale press:
.
The scale will display zero and following pictograms:
and
.
Zeroing is only possible within the scope of ±2% of full scale. While
zeroing outside the scope of ±2% you will see
possible only in stable state.
Notice:
Zeroing is possible only within the ±2% interval of the maximal range.
If zeroing is performed beyond this range the
short audible signal will be emitted.
11.4. Weighings in two ranges
Switching between the I range and the II range happens automatically
(exceeding Max of the I range). Weighings in the second range is
signalled by a pictogram in the top left corner of the display.
Then weighings is done with the accuracy of the II range to the moment
of returning to zero (autozero range
) where the scale switches back
to the I range.
- 26 -