Next Advance SP570-M User Manual
Page 35
NEXT>>>ADVANCE
www.nextadvance.com
SP500 series OEM Programmable Syringe Pump
NE500 series manual
30
Revision 101101
PUMP MOTOR OPERATING TTL OUTPUT CONFIGURATION
ROM [
Set/query Pump Motor Operating TTL output configuration (TTL pin 7)
Settings => 0 = Output is logic high only when the pump motor is operating (pumping).
1 = Output is logic high when the pump motor is operating (pumping) or when the
Pumping Program is executing a pause timer.
SET KEY BEEP
BP [
Set/query key beep mode. Set key beep enables or disables key beep mode.
This command is not relevant to models SP500 series pumps.
8.5.4 General Control and Status Commands
TTL I/O OUTPUT SETTING
OUT
Sets TTL level on user definable output pin on the ‘TTL I/O’ connector.
Valid value: 5 (Program Output pin)
Query is not applicable.
TTL INPUT QUERY
IN
Queries TTL level of pin on ‘TTL I/O’ connector. Set is not applicable.
Valid values: 2, 3, 4, and 6.
Response:
BUZZER
BUZ [ 0 | { 1 [ < n > ] } ]
Sets / queries buzzer
Set:
0 = Turn buzzer off;
1 = Turn buzzer on
if
If
otherwise buzzer beeps
if
Query response: { 0 | 1 }
0 = Buzzer off
1 = Buzzer is on continuously or beeping.
FIRMWARE VERSION QUERY
VER
Response: NE50
where ‘
Set is not applicable.