Next Advance SP570-M User Manual
Page 21

NEXT>>>ADVANCE
www.nextadvance.com
SP500 series OEM Programmable Syringe Pump
NE500 series manual
16
Revision 101101
Phase
Function
Rate
Volume
Direction
3
INC
1.0
0.1 ml
Infuse
Phase
Function
4
LOP 50
Phase
Function
5
LPS
Phase
Function
Rate
Volume
Direction
6
DEC
1.0
0.1 ml
Infuse
Phase
Function
7
L0P 99
Phase
Function
Rate
Volume
Direction
8
DEC
1.0
0.1 ml
Infuse
Phase
Function
9
LPS
Phase
Function
Rate
Volume
Direction
10
INC
1.0
0.1 ml
Infuse
Phase
Function
11
LOP 50
Phase
Function
12
JMP 02
7.4.4 Example 4: Complex Dispenses with External Synchronization
0
0
750
Syringe
Refill
A more complex dispensing example, this Program contains different pumping requirements, including
dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate
during the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the
pump waits for a start trigger before starting the next dispense.
The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set
of dispenses, the syringe is refilled by the amount infused, 17.25 ml. Then the buzzer beeps, to alert the
operator to the start of the first set of dispenses. The process is then repeated.
Phase
Function
Rate
Volume
Direction
1
RAT
750.0 ml/hr
0.5 ml
Infuse
Phase
Function
Rate
Volume
Direction
2
RAT
300.0 ml/hr
1.5 ml
Infuse