SP500 series OEM Programmable Syringe Pump
NE500 series manual
i
Revision 101101
Quick Start Instructions
Quick start instructions assume that the pump was not previous programmed with a multiple
Phase Pumping Program.
Plug in the pump.
Attach the computer cable to the "Computer" connection on the RJ-11 connector
Start the terminal emulator program on the computer
Baud Rate: 19,200
Data frame: 8N1
Note: Each command sent to the pump must have a carriage return (0x0D) at the end, but space
characters are optional and will be ignored.
Note: Wait for the response from each command before sending the next command
Setup Pumping Parameters
Set the Syringe Inside Diameter:
Where "nn.nn" is the syringe inside diameter: Send the command: DIA nn.nn
Set the Pumping Rate.
Where "nn.nn" is the pumping rate and "xx" is the rate's units: Send the command: RAT nn.nn xx
Rate units "xx" value
mL/hr
MH
µL/hr
UH
mL/mn
MM
µL/mn
UM
If the pumping rate is out of range the response message to the computer will include "?OOR"
Set the Volume to be Dispensed or Continuous Pumping
Where "nn.nn" is the volume to be dispensed, send the command: VOL nn.nn
If the volume to be dispenses is 0, then the pump will pump continuously.
Note: The units for the volume to be dispensed will be mL or µL, depending on the syringe size.
At or below 14.00 mm syringe diameter, volume units are µL
Above 14.00 mm syringe diameter, volume units are mL
Set the Pumping Direction
For infusion, send the command:
DIR INF
For Withdrawal, send the command:
DIR WDR
Load the Syringe
Press in the white drive-nut button to move the pusher block.
Insert the syringe plunger in the pusher block slot.
Insert the syringe barrel flange in the flange brackets with the syringe barrel holder on the syringe.
Tighten the flange brackets onto the syringe flange. Tighten the pusher block screw.
Operating the Pump
Start the Pump
Send the command: RUN
Stop the Pump
Send the command: STP
Query the Pump Status
Send just a carriage return to the pump. The pump will respond with a status prompt
When Pumping
The pumping rate can be changed.
With continuous pumping, the pumping direction can be changed.