Next Advance SP570-M User Manual
Page 32
NEXT>>>ADVANCE
www.nextadvance.com
SP500 series OEM Programmable Syringe Pump
NE500 series manual
27
Revision 101101
Rate Data Functions
When a Phase’s function is set to a “Rate Data Function”, use the ‘RAT’, ‘VOL’, and
‘DIR’ commands to setup the pumping parameters.
RAT
Pumping rate. ‘RATE’
INC
Increment rate. ‘INCR’
DEC
Decrement rate. ‘DECR’
Non-Rate Data Functions
STP
Stop pump. ‘STOP’
JMP
Jump to Program Phase.
LOP
Loop to previous loop start ‘nn’ times.
LPS
Loop starting Phase.
LPE
Loop end Phase.
PAS
Pauses pumping for ‘nn’ seconds.
PAS n.n
Pauses pumping for 'n.n' seconds.
IF
If Program input TTL pin low, jump to Phase.
EVN
Set event trigger trap.
EVS
Set event square wave trigger trap.
EVR
Event trigger reset.
BEP
Sound short beep.
OUT
Set programmable output pin.
PUMPING RATE
RAT [
Set/query pumping rate.
=
l/min
MM
= ml/min
UH
=
l/hr
MH
= ml/hr
Applicable only with “Rate Data Functions”.
When setting the pumping rate, if the current Phase’s function is not ‘RATE’, then
not applicable.
While pumping, the pumping rate can only be set if the current Phase function is ‘RATE’ and the
next Program Phase’s function to be executed is not ‘INCR’ or ‘DECR’. Also, while pumping,
The new pumping rate will only be stored in non-volatile memory if the Pumping Program is not
operating.
When the pumping rate is queried while pumping, the response will be the current pumping rate
and units. Otherwise, the response will be the rate setting and units, if applicable. With the
‘INCR’ and ‘DECR’ functions, these two responses are not the same.
VOLUME TO BE DISPENSED
VOL [
Set/query volume to be dispensed. Applicable only with “Rate Data Functions”. Can only be set
when the Pumping Program is not operating. The volume units are set according to the current
syringe diameter setting.
Query response: