SP500 series OEM Programmable Syringe Pump

NE500 series manual

27

Revision 101101

=>

Rate Data Functions

When a Phase’s function is set to a “Rate Data Function”, use the ‘RAT’, ‘VOL’, and
‘DIR’ commands to setup the pumping parameters.

RAT

Pumping rate. ‘RATE’

INC

Increment rate. ‘INCR’

DEC

Decrement rate. ‘DECR’

Non-Rate Data Functions

STP

Stop pump. ‘STOP’

JMP

Jump to Program Phase.

LOP

Loop to previous loop start ‘nn’ times.

LPS

Loop starting Phase.

LPE

Loop end Phase.

PAS

Pauses pumping for ‘nn’ seconds.

PAS n.n

Pauses pumping for 'n.n' seconds.

IF

If Program input TTL pin low, jump to Phase.

EVN

Set event trigger trap.

EVS

Set event square wave trigger trap.

EVR

Event trigger reset.

BEP

Sound short beep.

OUT

Set programmable output pin.

PUMPING RATE

RAT [ [ ] ]

Set/query pumping rate.

=> UM

=

l/min

MM

= ml/min

UH

=

l/hr

MH

= ml/hr

Applicable only with “Rate Data Functions”.

When setting the pumping rate, if the current Phase’s function is not ‘RATE’, then is
not applicable.

While pumping, the pumping rate can only be set if the current Phase function is ‘RATE’ and the
next Program Phase’s function to be executed is not ‘INCR’ or ‘DECR’. Also, while pumping,
can not be set.

The new pumping rate will only be stored in non-volatile memory if the Pumping Program is not
operating.

When the pumping rate is queried while pumping, the response will be the current pumping rate
and units. Otherwise, the response will be the rate setting and units, if applicable. With the
‘INCR’ and ‘DECR’ functions, these two responses are not the same.

VOLUME TO BE DISPENSED

VOL [ ]

Set/query volume to be dispensed. Applicable only with “Rate Data Functions”. Can only be set
when the Pumping Program is not operating. The volume units are set according to the current
syringe diameter setting.

Query response: