WIKA operating instructions calibration system, model BCS10

14051046.01 10/2012 GB/D/F/E

GB

8. Faults / 9. Dismounting, return and disposal

CAUTION!

If faults cannot be eliminated by means of the measures listed above,

shut down the calibration system immediately, and ensure that pressure

and/or signal are no longer present, and secure the instrument from being

put back into operation inadvertently.

In this case, contact the manufacturer.

If a return is needed, please follow the instructions given in chapter 9.2

"Return".

8.2 Test pump fault

Problem

Causes

Measures

The test pump is leaking

Pressure relief valve is open

The pressure and vacuum switch-

ing is not switched correctly

(The switching is in the middle

position)

Close the pressure relief valve

Turn the shift lever to pressure

generation.

The pump handles are stiff

to operate

The pump has not been used for a

long time

Actuate the pump until it

functions smoothly

CAUTION!

If faults cannot be eliminated by means of the measures listed above,

shut down the calibration system immediately, and ensure that pressure

and/or signal are no longer present, and secure the instrument from being

put back into operation inadvertently.

In this case, contact the manufacturer.

If a return is needed, please follow the instructions given in chapter 9.2

"Return".

9. Dismounting, return and disposal

WARNING!

Residual media on submersible calibration system can result in a risk to

persons, the environment and equipment.

Take sufficient precautionary measures.

9.1 Dismounting

Only dismount measurement setups once the system has been depressurized!