WIKA operating instructions calibration system, model BCS10

14051046.01 10/2012 GB/D/F/E

GB

3. Specifications / 4. Design and function

Voltage supply

Power supply:

3 x 1.5 V AA alkaline batteries

Battery life:

approx. 1.500 ... 2.000 h

Battery status indication: Icon in display for low battery level

Process connection

Test item adapter with quick-acting coupling for G ½

Test item adapter with quick-acting coupling for G ¾

Test item adapter with quick-acting coupling for M30 x 2

CE conformity

EMC directive:

2004/108/EC, EN 61 326-1 emission (group 1, class B)

and interference immunity (appendix A)

Plastic case

Dimensions in mm:

395 x 295 x 106

Weight:

approx. 4 kg (with contents)

For further specifications see WIKA data sheet SP 60.08 and the order documentation.

4. Design and function

4.1 Short description

The modular calibration system model BCS10 serves for pressure generation and

inspection of mechanical and electronic SF

6

gas density measuring instruments by

means of comparative measurements at 20 °C.

The test pump can generate a pressure of 0 ... 35 bar. Individual measuring points can

be approached with highest accuracy by means of the fine-adjustment valve.

The results of the measurements can be displayed in one of the 21 measuring units or in

customer units using the built-in digital pressure gauge.

4.2 Scope of delivery

Precision digital pressure gauge model CPG1000 with attached pneumatic test

pump model CPP30

Test item adapter with quick-acting coupling for G ½

Test item adapter with quick-acting coupling for G ¾

Test item adapter with quick-acting coupling for M30 x 2

Adjustable overpressure valve

Metal flexible hose for connecting the test item, length 2 m

Plastic case, incl. foam insert nad operating instructions

Cross-check scope of delivery with delivery note.