Gb 3. specifications / 4. design and function, Design and function – WIKA BCS10 User Manual
Page 11

11
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
GB
3. Specifications / 4. Design and function
Voltage supply
■
Power supply:
3 x 1.5 V AA alkaline batteries
■
Battery life:
approx. 1.500 ... 2.000 h
■
Battery status indication: Icon in display for low battery level
Process connection
■
Test item adapter with quick-acting coupling for G ½
■
Test item adapter with quick-acting coupling for G ¾
■
Test item adapter with quick-acting coupling for M30 x 2
CE conformity
■
EMC directive:
2004/108/EC, EN 61 326-1 emission (group 1, class B)
and interference immunity (appendix A)
Plastic case
■
Dimensions in mm:
395 x 295 x 106
■
Weight:
approx. 4 kg (with contents)
For further specifications see WIKA data sheet SP 60.08 and the order documentation.
4. Design and function
4.1 Short description
The modular calibration system model BCS10 serves for pressure generation and
inspection of mechanical and electronic SF
6
gas density measuring instruments by
means of comparative measurements at 20 °C.
The test pump can generate a pressure of 0 ... 35 bar. Individual measuring points can
be approached with highest accuracy by means of the fine-adjustment valve.
The results of the measurements can be displayed in one of the 21 measuring units or in
customer units using the built-in digital pressure gauge.
4.2 Scope of delivery
■
Precision digital pressure gauge model CPG1000 with attached pneumatic test
pump model CPP30
■
Test item adapter with quick-acting coupling for G ½
■
Test item adapter with quick-acting coupling for G ¾
■
Test item adapter with quick-acting coupling for M30 x 2
■
Adjustable overpressure valve
■
Metal flexible hose for connecting the test item, length 2 m
■
Plastic case, incl. foam insert nad operating instructions
Cross-check scope of delivery with delivery note.