WIKA operating instructions calibration system, model BCS10

14051046.01 10/2012 GB/D/F/E

GB

6. Commissioning, operation

6. Commissioning, operation

Commissioning may only be carried out by trained and qualified personnel. The calibra-

tion system was comprehensively checked ex works for tightness.

WARNING!

Only connect or disconnect test and calibration installations once the

system has been depressurised!

6.1 Switching on and off

To switch on the calibration system,

press the Power key (1). Press the

Power key again to switch off the

calibration system.

After switching on the version of the

firmware is indicated on the display for

approx. 1 second.

6.2 Zero point adjustment

To set the displayed value as zero

point, simply press the ZERO key (3).

The display changes immediately to the

0 value.

6.3 MIN/MAX value

The digital pressure gauge saves the

minimum and maximum pressure.

If the MAX/MIN key (4) is pressed once,

the maximum pressure from the memory is displayed. If the MAX/MIN key is pressed

again, the minimum pressure is displayed from the memory.

After 2 seconds, the display switches back to the operating mode. To clear the MAX/MIN

memory, press the MAX/MIN key (4) for at least 2 seconds until the message "CLr" is

displayed.

The bargraph display at the bottom of the display indicates the applied pressure relative

to the total measuring range.

Due to a possible Tare setting, the displayed pressure value can differ

from the applied pressure.