beautypg.com

Gb 3. specifications 3. specifications, Specifications – WIKA CEP1000 User Manual

Page 8

background image

8

WIKA operating instructions, model CEP1000

11570351.02 10/2010 GB/D/F/E/I

GB

3. Specifications

3. Specifications

Specifications

Input

Current range

0.000 … 24.000 mA

-25.00 … +125.00 %

Voltage range

DC 0.000 … 28.000 V

Input protection

Fuseless, up to AC 250 V

Output

Current range

0.000 … 24.000 mA

-25.00 … +125.00 %

Uncertainty

± 0.015 % of reading ± 2 µA

Resolution

1 µA, 1 mV

Drive capability

Without HART™-resistor

1200 Ω

With HART™-resistor

950 Ω

Loop supply voltage

DC 24 V

Range select

Decade; incremental to 0.001 mA steps

Special features

Auto step/ramp, Auto “%-Error” function, built-in resistor

for HART™ communications

Permissible

Operating temperature

-10 … +55 °C

Storage temperature

-20 … +70 °C

Display

5 1/2-digit with character size 9 mm and %-announce-

ment

Range

mA, V, %

Recording options

MIN-/MAX-storage

Power supply

DC 9 V battery

Battery life

12 hours (12 mA into 250 Ω continuous)

Low battery indicator

displayed icon near the end of battery life

EMC / CE conformity

2004/108/EC, EN 61326 Emission (Group 1, Class B) and

Immunity (portable equipment)

Ingress protection

IP 52

Calibration

3.1 calibration certificate per DIN EN 10204 (option: DKD/

DAkkS-calibration certificate)

Dimensions

144.7 x 80.0 x 36.3 mm

Weight

approx. 340 g

For further specifications see WIKA data sheet CT 81.01 and the order documentation.